FLUKE 1529 Chub-E4 Standards Thermometer
- Supports four channels for PRTs, thermistors, and thermocouples, accommodating a variety of sensor types.
- Display eight user-selected data fields from any channel.
- Log up to 8,000 readings with date and time stamps for data tracking.
- Eight hours of continuous operation
Related products
-
Temperature Calibrators
FLUKE Calibration 9142 Field Metrology Well
- Lightweight, portable, and fast
- Built-in two-channel readout for PRT, RTD, thermocouple, 4-20 mA current
- True reference thermometry with accuracy to ±0.01 °C
- On-board automation and documentation
- Metrology performance in accuracy, stability, uniformity, and loading
(0 reviews) -
Temperature Calibrators
FLUKE 9132/9133 Portable Infrared Calibrator with Blackbody Source
Product overview: Fluke 9132/9133 Portable Infrared Calibrator with Blackbody Source
Reliable, Portable Blackbody Calibration for IR Thermometers and Imagers
Fluke’s 9132 and 9133 infrared calibrators are designed to meet the precise needs of infrared thermometer and thermal imager users. These portable blackbody sources offer fast heating and cooling, stable temperatures, and radiometric accuracy—making them ideal for field and lab environments alike.
Fluke 9132 – High-Temperature Blackbody Source (Up to 500 °C)
- Designed for IR calibrations above ambient
- Provides a stable blackbody surface up to 500 °C (932 °F)
- Short heating and cooling times keep workflows efficient
- Radiometric calibration ensures measurement confidence across its entire range
Fluke 9133 – Low-Temperature Blackbody Source (–30 °C to 150 °C)
- Ideal for calibrating IR thermometers at low temperatures
- Reaches –30 °C (22 °F) in typical ambient conditions using solid-state cooling
- Convenient dry gas fitting on front bezel prevents ice buildup on the target
- Offers stable temperature control up to 160 °C (320 °F)
- Heats or cools from ambient to either extreme in approximately 15 minutes
What is a Blackbody Calibrator?
A blackbody calibrator (or blackbody source) is a device that emits thermal radiation with high emissivity, typically near 1.0. This makes it behave closely to an ideal blackbody, emitting radiation according to Planck’s law.
Blackbody calibrators are essential tools for ensuring the accuracy of:
- Infrared thermometers
- Thermal imagers
- Non-contact temperature measurement devices
By aiming a device under test at the precisely controlled blackbody target, users can verify and adjust readings to ensure accuracy and compliance with calibration standards.
Summary: Why Choose the 9132 or 9133?
Model Temperature Range Ideal For 9132 Up to 500 °C (932 °F) High-temperature IR calibration 9133 –30 °C to 160 °C (–22 °F to 320 °F) Low-temperature IR calibration with fast transitions Both units provide:
- Portable, stable blackbody sources
- Quick heating and cooling times
- High emissivity targets for accurate infrared calibration
- Confidence backed by Fluke’s precision and reliability
(0 reviews) -
Intrinsically Safe
FLUKE 1552a Stik Thermometer & Temperature Calibrator
- Intrinsically-safe, battery operated, portable reference thermometer
- Provides accuracy of ±0.05°C (±0.09°F) over full range
- Features intrinsically-safe operation (ATEX and IECEx compliant)
- Measures -80 °C to 300 °C ( -112 °F to 572 °F)
- Provides user-configurable temperature with Trend/Stability indicator
- Displays temperature in °C or °F
(0 reviews) -
Temperature Calibrators
FLUKE 712B RTD Calibrators
- Highly accurate, full-featured , easy-to-use single function RTD temperature calibrator
- Measures and simulates 13 different RTD types and resistance
- Measures 4 to 20 mA signals while simultaneously sourcing a temperature signal
- Comes with a magnetic hanging tool designed in for convenient hands-free operation
- Features configurable 0% and 100% source settings for quick 25% linearity checks
(0 reviews)